What an ASTM D6319 Test Report Tells a Canadian Glove Buyer, and What It Leaves Out
A dental office orders nitrile gloves four times a year, and each delivery arrives with a carton and, sometimes, a PDF. The carton states a size, a quantity and a lot number. The PDF states that the product meets ASTM D6319. Between those two documents sits the question a purchaser actually has to answer: what was measured, on which gloves, and what does the answer not cover?
Reading the report properly is worth an hour, because it changes what you put in the purchase order. A buyer who knows that tensile strength is measured before and after ageing asks a different question about shelf life than one who only sees a compliance statement. A buyer who knows that freedom from holes is a sampled test rather than an assurance understands why the sample size matters more than the pass mark.
Quick facts
- ASTM D6319 is the specification for nitrile examination gloves used in medical examinations and in diagnostic or therapeutic procedures, and it covers gloves that fit either hand, paired gloves and gloves by size.
- Its abstract lists tests for freedom from holes, tensile strength and elongation at break before and after ageing, powder amount, sterility where claimed, and a talc-free requirement on both surfaces.
- The related European standard for single-use medical gloves splits those obligations: EN 455-1 covers freedom from holes, and EN 455-3 covers biological evaluation.
- Sampling is governed by ISO 2859-1, which pairs a lot size and an inspection level to produce a sample size, and an AQL to produce an accept or reject decision. The 1999 edition was withdrawn in January 2026 and replaced by a 2026 edition.
- Canadian sales are governed by establishment and device licensing obligations, which are separate from any performance specification. CliniEco Medical is a licensed medical device establishment (MDEL #35334).
- A report issued for one lot describes that lot. It does not travel with the next shipment unless the supplier reissues it.
What does ASTM D6319 actually measure on a nitrile exam glove?
The specification treats a glove as a barrier with mechanical properties, and tests each of those separately. Freedom from holes is the barrier test: gloves are filled or immersed and inspected for leakage, which catches a failure that a visual check on the line will miss. Tensile strength and elongation at break are the mechanical tests, and they are run twice, once on new product and once after accelerated ageing, because a nitrile film that is strong on the day it is made can lose performance in a warm storeroom.
Powder is treated as a separate property. The abstract allows an approved lubricant for absorbable dusting powder, permits other lubricants only where safety and efficacy have already been established, and requires both the inside and outside surfaces to be free of talc. For a clinic with staff who react to glove dust, that clause is the part of the specification that matters, and it is also the part most likely to be summarised as a single word on a marketing sheet.
| What the report section covers | What it establishes | What a buyer still has to decide |
|---|---|---|
| Freedom from holes | Whether the sampled gloves met the watertight test at the stated AQL | Whether the sample size was large enough to represent the lot |
| Tensile strength before ageing | Strength of the film as delivered | Whether the strength suits the task, such as prolonged wet work |
| Tensile strength after ageing | How the film behaves after accelerated ageing | Whether your storage conditions are kinder or harsher than the test |
| Elongation at break | How far the film stretches before it breaks | Whether staff report tearing at the cuff during donning |
| Powder and talc statement | What lubricant is present and that talc is absent | Whether residue matters for your procedures or staff skin |
| Sterility, where claimed | The sterilisation status of the packed product | Whether the task needs a sterile glove at all |
Which numbers on the report correspond to what staff feel at the chair?
Two failures dominate complaints in a clinic, and both map onto a specific line of the report. The first is a glove that tears while being pulled on, which is an elongation problem as much as a strength problem: a film that resists force but does not stretch far enough will split at the cuff when a hand enters quickly. The second is a glove that feels different three months into storage, which is what the after-ageing test is designed to catch.

The link between report and experience is why the edition matters. When a supplier changes the film, the report changes with it, and the practical effect appears at the chair rather than in the storeroom. Buying the same product code does not by itself confirm the same glove; the lot and the specification edition are what identify the batch you are wearing.
That also explains why a glove that passes every test can still be unpopular. Fit, fingertip taper, cuff length and the way the film grips a dry instrument are not in the specification. They are the reason two clinics with the same procedures and the same hand sizes end up ordering different products, and the reason a trial box is worth more than a specification sheet when a supplier is new.
How does AQL sampling decide how much of a lot is inspected?
An acceptance quality limit is not a promise that no defective glove exists. It is the defect rate that the sampling plan is designed to accept, paired with a sample size drawn from the lot. ISO 2859-1 sets out the lot-by-lot schemes, indexed by AQL, and the tables convert a lot size and an inspection level into a code letter and then into a number of gloves to test.
This is the part of the report buyers skim, and it is the part that carries the most information. A supplier claiming a tight AQL on a small sample, or a report that states an AQL without stating the sample size or the acceptance and rejection numbers, cannot be checked. ISO 2859-4 addresses a related question, the assessment of a declared quality level, which is the statistical exercise you are performing when you ask whether a shipment really meets the level claimed on the carton.
For a clinic the practical use is narrower than the statistics. You are not going to resample a pallet. You are deciding whether the document in front of you is specific enough to be evidence, and the test is whether it names the lot, the sample size, the failures and the edition. A report that names all four can be filed next to the purchase record and used later in a recall question.
What does the report leave out?
Three gaps matter in practice. The first is chemical permeation. A standard nitrile examination report does not describe resistance to chemotherapy drugs, and a clinic that handles hazardous drugs needs permeation data for those specific drugs, requested as a separate document. The related standards for protective gloves against chemicals and against micro-organisms, ISO 374-1 and ISO 374-5, define how that performance is classified; they are not part of a D6319 examination report.
The second gap is biological evaluation. Biocompatibility is handled through ISO 10993-1 as part of a risk management process under ISO 14971, and the resulting statement is a manufacturer declaration rather than a laboratory line item on a glove report. If a clinic is buying for staff with known sensitisation, the useful question is not the tensile number but which accelerators and residues the supplier has documented.

The third gap is traceability inside your own building. The report ties performance to a lot, but nothing in the document tells you which storeroom shelf that lot sits on, or which treatment rooms it reached. That link is created by the receiving record, which is why the lot number belongs on the packing slip and in the supply log rather than only on the carton.
What documentation should accompany a case in Canada?
Four documents do the work, and they answer different questions. The test report answers whether the product met a specification. The quality management certificate, typically ISO 13485, answers whether the manufacturer runs a system that would catch a change. The label answers what is inside the box, and ISO 15223-1 defines the symbols used for lot number, expiry, single use and sterile status so a receiving clerk can read a carton printed in another language.
The fourth is the supply agreement itself, because the performance claim only means something if the contract names the specification and the edition. A purchase order that says nitrile gloves without naming D6319 or EN 455 leaves the supplier free to change specification between shipments without telling you.
| Document | Question it answers | Who should hold it |
|---|---|---|
| Test report for the lot | Did this lot meet the named specification | Purchaser, filed with the receipt |
| ISO 13485 certificate | Is there a quality system behind the production | Purchaser, current copy |
| Label per ISO 15223-1 | What is in the box, when does it expire, is it sterile | Receiving, on the carton |
| Safety data sheet, where applicable | What is on the glove and how to handle a spill | Health and safety lead |
| Purchase order naming the specification | What the supplier agreed to supply | Purchaser, retained |
Products in a clinic order rarely share one evidence set, which is why the case lines are worth separating. A standard 3.5 mil blue nitrile glove covers routine examination work, and [the 1,000-glove case of 3.5 mil standard blue nitrile gloves](https://clinieco.ca/products/clinieco-medical-nitrile-gloves-3-5mil-standard-blue-1000-case) is the line most clinics reorder without discussion. A [4.0 mil chemo-rated Protégé glove](https://clinieco.ca/products/aurelia-nitrile-examination-gloves-protege-4-0-mil-powder-free-chemo-grade-100-count) is a different purchase, because the permeation data behind it is a different document. A [6 mil heavy-duty black glove](https://clinieco.ca/products/clinieco-medical-nitrile-gloves-6mil-heavy-duty-black-100-pack) is bought for mechanical work such as instrument transport and cleaning, where film strength matters more than tactile sensitivity. Mixed pallets are common, and [the wholesale case configuration of 10 boxes of 100](https://clinieco.ca/products/intco-nitrile-gloves-wholesale-case-touch-flex-advancare-10-boxes-100-ct) is how most multi-room practices keep sizes together without ordering four separate products.
What should a buyer request before a first order?
Ask for five items and read them in order. The report for the lot being shipped, with sample size and acceptance numbers. The specification and edition named in the purchase order. The quality management certificate and its expiry. The label artwork, so receiving can match symbols to the carton. And the shelf life statement, so the order quantity matches the rate at which the practice actually consumes the product.
Then check one thing that no document can answer. Staff handling a full day of examinations will tell you within a week whether the film suits their hands, and a trial box costs less than a quarter of documentation review. Buyers who skip that step tend to discover a fit problem after the second case has arrived.
For practices consolidating glove purchasing across rooms, the [glove collection](https://clinieco.ca/collections/gloves) groups the sizes and thicknesses in one place, and the [family practice clinic supply page](https://clinieco.ca/pages/family-practice-clinic) sets out the consumables that usually travel with a glove order. Orders that need lot-level documentation and a single invoice for several rooms can be arranged through the [wholesale and bulk quote desk](https://clinieco.ca/pages/wholesale), and clinics that keep receiving records for assessment can log the lot and receipt date in the [compliance log centre](https://clinieco.ca/pages/compliance-log-center). Reselling? become a distributor.
The habit that makes all of this manageable is small: file the report with the packing slip, write the lot number on the receiving record, and note the specification edition in the purchase order. Three fields, once per delivery, and the recall question that takes an afternoon to answer becomes a two-minute lookup.
References
- ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application (checked 9 November 2026)
- EN 455-1:2000, Medical gloves for single use, requirements and testing for freedom from holes (checked 9 November 2026)
- EN 455-3:2015, Medical gloves for single use, requirements and testing for biological evaluation (checked 9 November 2026)
- ISO 21420:2020, Protective gloves, general requirements and test methods (checked 9 November 2026)
- ISO 374-1:2016, Protective gloves against dangerous chemicals, terminology and performance requirements (checked 9 November 2026)
- ISO 374-5:2016, Protective gloves against micro-organisms (checked 9 November 2026)
- ISO 2859-1:1999, Sampling procedures for inspection by attributes, schemes indexed by acceptance quality limit (checked 9 November 2026)
- ISO 2859-4:2020, Sampling procedures for inspection by attributes, assessment of declared quality levels (checked 9 November 2026)
- ISO 10993-1:2018, Biological evaluation of medical devices, evaluation and testing within a risk management process (checked 9 November 2026)
- ISO 14971:2019, Medical devices, application of risk management to medical devices (checked 9 November 2026)
- ISO 13485:2016, Medical devices, quality management systems (checked 9 November 2026)
- ISO 15223-1:2021, Symbols to be used with information supplied by the manufacturer (checked 9 November 2026)
- OSHA 1910.1030, Bloodborne Pathogens (checked 9 November 2026)
- CCOHS, Gloves, OSH Answers (checked 9 November 2026)
- Public Health Ontario, Reprocessing and cleaning of patient care equipment (checked 9 November 2026)
- World Health Organization, WHO Guidelines on Hand Hygiene in Health Care (checked 9 November 2026)
Printable companion: the free sterilization log sheet lays out the fields an inspection asks for, so the record is written down as you work instead of reconstructed later.
Want the sheet filled in for your cycle type rather than a blank form? The sterilization log generator prints a record with the columns your reprocessing documentation needs.
Buying for a facility? Open a wholesale and institutional account for case pricing, purchase-order billing and a named account contact.
Related Reading
- What Does Glove Cost per Patient Visit Run in a Canadian Clinic, and Where Does It Leak?
- Why Nitrile Gloves Fail Before the Expiry Date: Heat, Ozone and Tensile Retention
Frequently Asked Questions
What is an ASTM D6319 test report?
It is the laboratory summary behind ASTM D6319, the specification for nitrile examination gloves used in medical examinations and diagnostic or therapeutic procedures. It lists the tests performed, the results, the lot or batch sampled and the standard edition applied. It is a supplier document, not a Canadian licence.
Does a D6319 report prove a glove is legal to sell in Canada?
No. The report speaks to performance against a specification. Selling in Canada depends on medical device establishment licensing and device licensing obligations that sit with the licence holder, which is a separate set of paperwork from the laboratory summary.
Which tests appear most often on the report?
Freedom from holes, tensile strength and elongation at break before and after accelerated ageing, and powder residue on the inside of the glove. Sterility is added when the product is sold sterile rather than bulk non-sterile.
What is the AQL on a glove report?
The acceptance quality limit is the defect rate the sampling plan is designed to accept. It is paired with a sample size drawn from the lot, so a report that states an AQL without the sample size and the number of failures cannot be checked by the buyer.
Does the report cover chemotherapy handling?
Only if the glove was tested for chemotherapy drug permeation and the report says so. A standard examination glove report describes barrier and strength tests, and a clinic handling hazardous drugs should ask for the permeation data separately.
Can a clinic ask for the report before ordering?
Yes, and the request should name the lot. A certificate issued for a different production run describes a different set of gloves, and the value of the document is that it follows the boxes actually delivered.
How long should a clinic keep glove documentation?
Keep the report, the purchase record and the lot number for as long as the boxes could still be in use, then for the retention period your accreditation programme expects for purchasing and quality records. The lot number is what connects a recall notice to your storeroom.
What does a report not tell you about fit?
Nothing about fit. Length, width, fingertip taper and cuff behaviour are physical properties that staff feel, and a glove that passes every laboratory test can still be the wrong shape for the hands that wear it for six hours.
Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).