How Many Class 5 Chemical Integrators Does a Two-Operatory Hygiene Clinic in Ontario Use in a Month?
A hygiene clinic does not buy Class 5 integrators by the load. It buys them by the package, because the indicator goes inside the package the patient's instruments are wrapped in. Once that is settled, the monthly count stops being a guess.
The strip in question is the Class 5 steam chemical integrator, ISO 11140-1 Type 5, sold as a 200-strip pack and as a case of five. It reads yellow to black through accept and reject windows across 121 to 135 C, which lets a hygiene clinic read a load at the end of the cycle instead of waiting for a spore result.
How many Class 5 integrators does one sterilization package need?
One. Public Health Ontario's clinical office practice guidance is explicit that an internal chemical indicator is placed inside each package, container or bundle that undergoes sterilization, and that does not change whether the package is a pouch or a wrapped tray. A Type 5 integrator satisfies it and adds what a Class 1 tape cannot: the colour change is tied to time, temperature and saturated steam together rather than to temperature alone.
How many does a two-operatory hygiene clinic use in a month?

Two hygiene operatories at eight appointments each, with 1.5 packages per appointment for the cassette plus a pouch for the mirror or prophy tip, gives 24 packages a day: 504 in 21 working days, and 504 strips.
The table scales it, and the last two columns are why pack size matters more than pack price. Those packages are sealed in Class 4 self-sealing pouches.
| Operatories | Packages a day | Strips a month | 100-strip packs | 200-strip packs |
|---|---|---|---|---|
| 1 | 12 | 252 | 2.5 | 1.3 |
| 2 | 24 | 504 | 5.0 | 2.5 |
| 4 | 48 | 1,008 | 10.1 | 5.0 |
Why is the 100-per-pack figure wrong for this order?

Because the pack sells in 200s. The Type 5 line is listed and priced as a 200-strip pack and a case of five, so planning against the older 100-strip wording doubles the pack count. At 504 strips a month a 200-strip pack covers about eight and a half working days.
What pushes the monthly count up or down?
Three variables, and only one is patient volume. Package discipline comes first: a clinic that wraps instruments individually instead of using a cassette turns one package into four, and the strip count follows. Unused packages come second, since a package opened without being used consumes its indicator anyway. The load-release device is third, a per-cycle cost that does not shrink on a light schedule.
Cycle length does not move the count, which is why the integrator line scales with appointment volume rather than with sterilizer settings. The 24-hour biological indicator, 25 tests to a pack is the other half of that pair and scales with days instead.
Running a dental practice? See our dental practice sterilization and hygiene supplies.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
References
- Public Health Ontario, Infection Prevention and Control for Clinical Office Practice, November 2026 (checked 16 November 2026)
- ISO 11140-1:2014, Sterilization of health care products, chemical indicators (checked 16 November 2026)
- CSA Z314:23, Canadian medical device reprocessing in all health care settings (checked 16 November 2026)
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices (checked 16 November 2026)
Printable companion: the compliance log center carries the indicator and load log that pairs the package count with the strip used, which is what an inspection reviews when it wants evidence that every package carried an internal indicator.
Counting packages rather than instruments? The clinic supply planner keeps the package estimate per operatory on its own row so the strip order follows the schedule rather than the last invoice.
Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you’ll get a written answer specific to your equipment and province, with the regulation or standard cited.
Related Reading
- Class 5 Steam Chemical Integrators: Reading the Window, Pack Counts and Load Release
- Autoclave Tape vs Chemical Integrators: Which One Actually Verifies Sterility
Frequently Asked Questions
Can a Class 5 integrator replace the daily biological indicator?
No. The integrator shows whether the cycle reached sterilizing conditions inside the package; the biological indicator shows whether the spores were killed. Ontario offices run both.
Is there a 100-strip pack of Class 5 integrators?
The Type 5 line ships a 200-strip pack and a case of five packs. Buying guides still quote a 100-strip pack, so a budget written on that figure plans twice the packs a 200-strip order needs.
How many does a two-operatory hygiene clinic use in a month?
About 500. At 24 packages a working day across 21 working days the clinic runs 504 packages, and one strip per package is 504 strips.
How many Class 5 integrators go in one sterilization package?
One, when the package carries no internal indicator of its own. Ontario guidance requires an internal chemical indicator in each package, and a Type 5 integrator meets that while giving a read tied to the biological result.
Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).