Record the manufacturer, catalogue number, lot number, expiry date, receiving date, acceptance-test result and the storage location, and then record which lot was in use for each analytical run or batch of specimens. ISO 15189:2022 puts acceptance testing and inventory control of reagents and consumables in clause 6.6, and puts the records themselves in clause 8.4.
The reason a laboratory bothers is not the tube. It is the recall. A centrifuge tube that cracks under load, a container whose cap leaks in transport, a lot of tubes whose graduations are misprinted — each of those becomes a question about which specimens were affected, and the answer exists only if the lot number was written down at the point of use rather than reconstructed afterwards from a purchase order.
The consumables in that chain are ordinary bench stock: the graduated leak-proof specimen cups, 100 per pack are the container most Canadian clinics and small laboratories order, and the lot printed on the carton is the identifier the whole record hangs on.
Quick facts
- ISO 15189:2022 places reagents and consumables in clause 6.6, which covers receiving and storage, acceptance testing (6.6.3), inventory management (6.6.4), instructions for use (6.6.5) and adverse event reporting (6.6.6).
- The same standard's clause 8.4 governs control of records, including how long they are kept — a period the laboratory sets against national, regional and local requirements rather than a single international number.
- Inventory management under 6.6.4 requires accepted consumables to be kept separate from items not yet checked or not accepted for use, which is what makes an acceptance-test record operationally meaningful.
- In the United States, CLIA 42 CFR 493.1105 requires analytic systems records — including quality control and patient test records and instrument printouts — to be retained for at least two years, with longer periods for pathology and cytology material.
- Canadian laboratories are accredited through programmes such as IQMH in Ontario and Accreditation Canada, both of which assess against ISO 15189.
- A lot number that is recorded only on the purchase order, and not at the bench, cannot answer a recall question about which specimens were collected in the affected containers.
Which lot fields have to be recorded when containers arrive?
The arrival record is the base of the trace. It is also the record most often reduced to a delivery note.
| Field | Content | Why it matters |
|---|---|---|
| Manufacturer and catalogue number | Name and product code | Identifies the exact product, not just the category |
| Lot or batch number | The manufacturer's lot, exactly as printed | The identifier a recall notice will quote |
| Expiry date | As printed, with the format recorded | Expired stock must not reach the bench |
| Quantity received | Units or cases | Reconciles the delivery against the order |
| Receiving date | Date of receipt at the laboratory | Starts the traceability clock |
| Storage location | Room, shelf, or quarantine area | Shows the item was stored as required |
| Acceptance test result | Pass or fail, with the test performed | The decision that releases the lot to use |
| Acceptance tester | Name or initials | Accountability for the release decision |
| Certificate or declaration | Sterility, RNase-free or other claim as applicable | Supports any claim the laboratory relies on |
The lot number is the field that has to be transcribed rather than photographed. A photograph of a carton in a shared drive is not a record until someone has read the number into the inventory system, because a recall search runs on text.
How is the lot tied to a patient result?
Through the run record. The tube lot and the container lot are recorded at the level at which they are consumed:
- Analytical runs. The run sheet or the laboratory information system record for a batch carries the reagent lots and the consumable lots in use, alongside the quality control results for that run.
- Specimen receipt. The accessioning record captures the container type and, where the laboratory's procedure requires it, the container lot.
- Problem investigations. When a tube fails — a crack during centrifugation, a cap that will not seal, a gel barrier that migrates — the investigation starts from the lot, because the first question is whether the failure is one unit or the whole batch.
- Recall response. A manufacturer's recall notice names a lot; the laboratory searches the inventory record for the lot and the run records for the period it was in use.
Two design choices make this workable. The first is that the field exists in the laboratory information system, so that the lot is captured at the bench rather than on a paper slip that gets filed. The second is that a lot change is an event — the quality control that accompanies a new lot is recorded with the lot number, which is what ISO 15189:2022 clause 6.6.3 asks for when it requires acceptance testing of consumables that can affect examination quality.
Do Canadian and American laboratories keep the same lot records?
The principle is shared; the citation and the retention period are not.
| Question | Canada | United States |
|---|---|---|
| Quality standard | ISO 15189:2022, assessed by IQMH, Accreditation Canada and other accreditation bodies | CLIA regulations, with CAP accreditation as the common additional layer |
| Consumable acceptance | Required where the consumable can affect examination quality (ISO 15189:2022, 6.6.3) | Required by the laboratory's own procedure under CLIA quality system requirements |
| Inventory segregation | Accepted stock separated from unchecked or rejected stock (6.6.4) | Set by laboratory policy under CLIA and accreditation standards |
| Retention period for patient test records | Set by the laboratory against national, regional and local requirements (8.4) | At least two years for analytic systems records under 42 CFR 493.1105, longer for pathology and cytology |
| Recall response | Driven by the accredited quality management system | Driven by CLIA and accreditation requirements plus the laboratory's procedure |
| Where the requirement is inspected | Accreditation assessment against ISO 15189 | State survey or accreditation survey against CLIA |
The difference that catches laboratories operating on both sides of the border is retention. A Canadian laboratory that adopts a two-year retention habit because its American parent does will find that its own accreditation body expects a period set against Canadian requirements — and the reasoning for the period has to be written down, not assumed.
What does acceptance testing of a new lot involve?
It depends on what the consumable can affect, and that is the test the standard applies. For tubes and containers, the checks that matter are physical and functional:
| Check | What is examined | Failure mode it catches |
|---|---|---|
| Visual inspection | Cracks, deformation, foreign matter, label legibility | Damaged units that will fail in the centrifuge or in transport |
| Dimensional check | Fit in the rotor, cap thread engagement | A tube that does not seat, or a cap that cross-threads |
| Volume check | Graduations against a measured volume | Misprinted graduations that change a dilution |
| Leak test | Cap and seal under the conditions of use | Leakage in transport or during centrifugation |
| Sterility claim verification | Certificate or declaration where sterility is relied on | An unverified claim |
| Performance check | Behaviour in the actual workflow — centrifugation, transport, storage | Failures specific to the laboratory's process |
The judgement the standard asks for is proportionate: a consumable that cannot affect examination quality does not need the same acceptance effort as one that does, but a tube that changes the volume of a sample or a container that leaks in transport plainly can.
What happens when a lot is defective?
Four steps, in this order:
- Segregate. Move the lot out of the usable stock and into a clearly marked hold area, so that no one draws from it while the investigation runs. ISO 15189:2022's inventory requirement that accepted stock be separated from unchecked stock exists precisely to make this possible.
- Assess the impact. Identify the period the lot was in use, the runs and specimens it touched, and whether results need review. This is where the run record earns its keep.
- Report. Where the defect is a problem with the product itself, report it to the manufacturer and, where the national scheme requires it, to the regulator; clause 6.6.6 covers adverse event reporting for reagents and consumables.
- Close the loop. Record the investigation, the disposition and the corrective action, and check whether the acceptance test that released the lot should have caught it.
The fourth step is the one that turns a complaint into a quality improvement, and it is the one an assessor will ask about.
Which standard and which clause?
The requirements sit in ISO 15189:2022, Medical laboratories — Requirements for quality and competence, which replaced the 2012 edition and restructured the requirements into resource, process and management-system clauses. Clauses 6.6.3 and 6.6.4 are the ones a laboratory cites when asked why a new lot of tubes needed a check and why the rejected stock sits behind a label. Clause 8.4 is the one to cite for the records themselves.
The safety dimension of handling, storing and disposing of the materials is covered separately by ISO 15190:2020, and the broader document set for general laboratory practice — including specimen collection and handling documents — is published by CLSI. The same lot discipline applies to the biological indicators a laboratory runs in its sterilizers, which are specified by ISO 11138-3:2017 and recorded by lot for exactly the same reason.
In Canada, accreditation against ISO 15189 is delivered by bodies such as IQMH in Ontario and Accreditation Canada nationally. In the United States, the record retention floor is set by 42 CFR 493.1105, which lists the analytic systems records that must be kept for at least two years and the longer periods that apply to pathology and cytology material.
Related reading
- What does ISO 15189 require for a Canadian medical laboratory?
- ISO 15189 vs CSA Z314: which sterilization records a Canadian medical laboratory must keep
- Receiving sterilization consumables at an independent laboratory: lot numbers, quarantine and release
Build the record before the assessment
The compliance log centre carries printable receiving and acceptance log sheets, and the log generator builds the columns for a laboratory that receives consumables in cases rather than units. If your lot record needs to be checked against ISO 15189 expectations, the compliance consultation desk is the route; institutional accounts and standing orders go through the account enquiry page. Reselling? become a distributor.
CliniEco Medical distributes laboratory consumables in Canada under MDEL #35334. This page summarises requirements published by ISO, CLSI, IQMH, Accreditation Canada and the United States federal regulations; the requirement that binds a specific laboratory is the one in its own quality manual and its accreditation programme.
Frequently Asked Questions
Is a lot number required on the patient report?
Not on the report itself. The requirement is that the laboratory can trace the consumable lot used for the examination, which is normally done through the run record and the inventory record rather than through the report sent to the clinician.
Do we need to record the lot number of every tube we use?
Record the lot at the level at which it is consumed — per run, per batch, or per accession, according to the laboratory's procedure. What must be possible is the reverse lookup: given a lot, identify the specimens and runs it touched.
How long should the lot records be kept?
As long as the records they support. Since the patient test record has a defined retention period, the lot record for the consumables used in that testing has to survive at least as long, or the trace breaks at the point where it is most likely to be needed.
Can the certificate of analysis replace acceptance testing?
No. A certificate is a manufacturer's declaration about the lot; acceptance testing is the laboratory's own check that the lot performs in its process. A laboratory can decide that the certificate reduces what it needs to test, but the decision and its rationale belong in the record.
What if the same lot is used in two different laboratories on one site?
Each laboratory's quality system records its own use. A shared inventory system can hold the receiving and acceptance record once, but the run-level trace belongs to the laboratory that performed the examination.
Does a sterile claim on a container need to be verified?
Where the laboratory relies on the claim, it needs a basis for doing so: a certificate, a declaration of conformity, or a specification in the product documentation. An unverified claim in a laboratory procedure is a gap an assessor will find, because the procedure says the laboratory relies on it.