A Canadian maker of Class I devices — a pouch line, an indicator line, a small accessory bought by dental practices and care homes — takes an order from a clinic in the United States and discovers that the phrase "we are Class I in Canada" does not travel. The two countries regulate different acts. Canada regulates sale and import inside Canada. The United States regulates the establishment that makes the device and the listing of the device itself, whoever makes it and wherever they sit.


This is a clause-level reading of both sides, followed by the part that causes the most avoidable delay: the reprocessing accessory path. Pouches, indicators and cleaning chemistry raise a classification question of their own, separate from the sterilizer's and separate from the Canadian file. The ranges that usually open this route are the Class 4 self-sealing sterilization pouches and the spore tests packed alongside them. Where a step depends on your specific product code, this article says so instead of guessing.
What does Canada require before the device leaves the country?
Start with the classification, because it decides which Canadian obligations attach. Section 6 of the Medical Devices Regulations establishes four classes, with Class I as the lowest risk, and section 7 provides that a device falling into more than one class is treated as belonging to the higher one. The rules themselves are in Schedule 1.
Class I is then the quiet class. Section 26 prohibits importing or selling a Class II, III or IV device unless the manufacturer holds a licence for it — Class I devices need no device licence at all. What Class I does not escape is the establishment chain. Section 44(1) provides that no person shall import or sell a medical device unless the person holds an establishment licence, and subsection (2) sets out who is outside that rule: a retailer, a health care facility, the manufacturer of a Class II, III or IV device, and, for a Class I device, a manufacturer that imports or distributes solely through a person who holds an establishment licence. Subsection 44(3) puts the matching duty on the importer, who must ensure that the person they import from holds an establishment licence.
Health Canada's guidance on medical device establishment licensing puts the scope plainly: the establishment licence extends to manufacturers of the lowest-risk class as well as to importers and distributors of devices in every class, and it is what permits a device to be imported or distributed in Canada. For a Class I manufacturer selling only outside Canada, that chain is not engaged — the licence protects the Canadian distribution route, not the foreign one.
Export itself appears in Part 4 of the Regulations. Section 89 states that, for the purposes of section 37 of the Act, Schedule 3 sets out the form to be used for an export certificate, and specifies who signs it: for a corporation, the senior executive officer in Canada, the regulatory officer in Canada, or an authorized agent; for an individual, the exporter or their authorized agent. Section 90 prohibits signing a certificate that is false or misleading or that omits something affecting its accuracy, and section 91 requires the exporter to keep completed export certificates at its principal place of business in Canada and to produce them to an inspector on request.
Read those together and the Canadian position is narrower than it first appears. Canada does not issue a permission to export a Class I device. It regulates the domestic sale and import chain, and it provides an export certificate of Canadian regulatory status for destinations that ask for one.
What does the United States require before the device arrives?
On the American side the unit of regulation is the establishment and the device listing, not the Canadian class. Section 807.20(a) requires the owner or operator of an establishment engaged in the manufacture, preparation, propagation, compounding, assembly or processing of a device intended for human use to register and to submit listing information, and it names the activities that catch businesses which do not think of themselves as manufacturers: initiating or developing specifications for a device made by a second party, repackaging or relabelling, and acting as an initial importer.
For a Canadian establishment the operative provision is 807.40. Any establishment in a foreign country engaged in manufacturing a device that is imported or offered for import into the United States must register and list its devices in FDA's electronic system, and an official correspondent facilitates communication between the establishment and FDA. Paragraph (b) requires the name, address and phone number of a United States agent as part of the registration; only one agent may be designated; the agent must reside or maintain a place of business in the United States; on FDA's request the agent assists with communication, answers questions about the products offered for import and helps schedule inspections; and changes in the agent's details must be reported within 10 business days. Paragraph (c) states the consequence that matters at the border: no device may be imported unless it is the subject of a device listing and is made at a registered establishment.
Section 807.41 adds the supply-chain disclosure. The foreign establishment must identify each known importer and each person who imports or offers for import its devices — including agents and brokers — and supply the premarket submission number and other identifying information for each.
The exemptions in 21 CFR 807.65 are worth reading carefully, because they are exemptions for classes of persons, not for classes of device. Raw material and component makers, makers of general purpose articles not labelled or promoted for medical use, retail establishments making final delivery to the end user, and businesses whose main role is dispensing or service are outside registration. Being Class I is not on that list. A small manufacturer still registers.
Premarket notification is a separate question. Section 807.81(a)(3) requires a 510(k) at least 90 days before introducing a device that is about to be significantly changed or modified in design, components, method of manufacture or intended use. Section 880.9 then limits every exemption: a device intended for a use different from that of a legally marketed device of the same generic type, or operating on a different fundamental scientific technology, still needs a premarket notification.
Where do pouches, indicators and cleaning chemistry sit?
This is where Canadian and American thinking diverge most often, because a clinic buys a pouch and an indicator as consumables while a regulator sees two devices with their own classifications.
Under 21 CFR 880.6850, a sterilization wrap includes a pack, sterilization wrapper, bag or accessories, intended to enclose another medical device that is to be sterilized and to maintain sterility until use; it is a Class II device. Under 21 CFR 880.2800, a 24-hour self-contained biological indicator carries a known number of microorganisms of known resistance on a carrier, and a physical or chemical indicator monitors one or more parameters of the process; both are Class II devices. The sterilizer's own file does not extend to either of them.
Cleaning chemistry runs on a different track. Section 807.65(c) exempts makers of general purpose articles, such as chemical reagents or laboratory equipment, whose uses are generally known to trained users and which are not labelled or promoted for medical uses. A cleaner promoted for reprocessing medical devices does not fit that description, so the labelling decides the position, and the classification question has to be answered on what the label and promotional material actually claim.
| Product | Regulation | Class | What the regulation covers |
|---|---|---|---|
| Sterilization pouch, wrap or bag | 21 CFR 880.6850 | Class II | A pack, sterilization wrapper, bag or accessory intended to enclose a device to be sterilized and to maintain sterility until use |
| Biological sterilization process indicator | 21 CFR 880.2800(a) | Class II | A known number of microorganisms of known resistance on a carrier, used to monitor the adequacy of the process |
| Chemical process indicator, including indicator tape | 21 CFR 880.2800(b) | Class II | A device that monitors one or more parameters of the process through a visible change |
| General purpose cleaning chemical | Outside device registration where 807.65(c) applies | Not a device on that basis | Reagents whose uses are generally known to trained users and which are not labelled or promoted for medical uses |
The two rows that decide whether a first shipment can move are establishment authorisation and product listing; the rest of the table exists to stop a team from assuming that a Canadian duty discharges an American one.
| Obligation | Canada — Medical Devices Regulations, SOR/98-282 | United States — 21 CFR |
|---|---|---|
| Premarket review of the device | Device licence for Class II, III and IV (section 26); none for Class I | 510(k) unless the device type is exempt (807.81); exemption limits at 880.9 |
| Establishment authorisation | Establishment licence to import or sell (44(1)), with the exceptions in 44(2); it extends to manufacturers of the lowest-risk class and to importers and distributors of every class | Annual establishment registration by the establishment that makes the device, including foreign establishments (807.20, 807.40) |
| Product listing | No separate listing; the device licence identifies the device for Class II to IV | Device listing for devices in commercial distribution, by the registering establishment (807.20, 807.40) |
| In-country representative | No equivalent requirement for a foreign buyer | One United States agent, resident or with a place of business in the US (807.40(b)) |
| Quality management system | Class I manufacturers must meet the applicable requirements of sections 10 to 20; no licence review | Part 820 QMSR: ISO 13485:2016 as incorporated, plus 820.10 additions; applies to accessories as finished devices |
| Labelling | Sections 21 to 23 of the Regulations | Part 801; unique device identification under Part 830 for devices not exempt |
| Problem reporting | Mandatory problem reporting in the Regulations for manufacturers and licence holders | Medical device reporting under Part 803 |
| Export paperwork | Export certificate under Part 4 (sections 89 to 91), signed and kept on file for an inspector | No reciprocal paperwork; the importer of record clears the shipment against listing and registration |
What does the quality management system regulation change?
Since 2 February 2026, Part 820 is the Quality Management System Regulation. Section 820.1(a)(1) applies the part to finished devices manufactured, imported or offered for import into the United States, while paragraph (a)(2) keeps component manufacturers outside it. Section 820.10 then requires a documented quality management system complying with ISO 13485:2016 as incorporated by reference, together with the additional requirements the section lists: unique device identification tied to Clause 7.5.8, traceability procedures tied to Clause 7.5.9.1 where Part 821 applies, complaint notification tied to Clause 8.2.3, and advisory notices handled under Part 806.
Design and development controls are not universal. They apply to Class II and Class III devices and to the Class I devices listed in 820.10(c) — devices automated with computer software, plus five named device types. Traceability for implantable devices applies to devices that support or sustain life. FDA's own description of the regulation notes that accessories are finished devices, so an accessory maker is inside the QMSR.
What does a first shipment look like, step by step?
- Classify the product in the American system first. Find the regulation that describes what the product does, not what it is called. A pouch is not a sterilizer accessory in the regulatory sense.
- Decide the premarket question. Check whether the device type is exempt from 510(k), then check the limits on that exemption in the part that contains the regulation.
- Appoint one United States agent. Confirm in writing who holds the role, that the address is a US place of business, and who tells FDA within 10 business days if it changes.
- Register the foreign establishment and list each device. Include the accessories as separate listings rather than as part of a kit entry.
- Identify the importers. Section 807.41 asks for each known importer and each person who imports or offers for import, with premarket numbers.
- Close the QMS gap on paper. Map your ISO 13485 records to 820.10's additions, especially UDI, complaint notification and advisory notices.
- Keep the Canadian file intact. Export certificates are signed, dated and retained at the principal place of business in Canada, and section 90 makes an inaccurate certificate an offence in its own right.
What are the five misreads that cost the most time?
- "Class I in Canada means exempt in the United States." The two classifications are independent. Exemption from 510(k) is a property of the American device type, not of the Canadian class.
- "Our MDEL covers American distribution." An establishment licence authorises import or distribution in Canada. It says nothing to FDA.
- "The sterilizer's clearance covers our pouches and indicators." Both sit under their own Class II regulations, 880.6850 and 880.2800.
- "The export certificate is the American requirement." Part 4 serves destinations that ask for proof of Canadian status. FDA's requirements are registration, listing and, where applicable, premarket notification.
- "Listing is approval." Listing is an inventory of what an establishment makes. It is not a finding of safety or effectiveness, and it does not substitute for a 510(k) where one is required.
None of this is exotic, and most of it is a documentation exercise rather than a redesign. The failure mode is sequence: teams treat the American filing as an export formality to be handled after the first order, when registration and listing have to exist before the device can be imported at all.
Where does this take a clinic or a supplier next?
For the practice on the receiving end, the documents that matter are the ones filed with the shipment: a printable compliance log sheet that records the sterilizer, the indicator lot and the result; the clinic supply planner for quantities across a quarter; and, where a classification or labelling question needs an answer in writing, ask your compliance question. Buyers ordering across several sites can review wholesale and multi-site ordering for pack sizes and lead times. Reselling? become a distributor.
Related reading
- CSA Z314 compared with CDC sterilization rules in Canada and the US
- Reprocessing ultrasound probes and endoscopes: Canada compared with FDA and CDC
- PHAC and CDC infection control guidance compared
- Seal integrity testing: F1929, F88, F1140, F2096 and ISO 11607-1
Frequently Asked Questions
Does a Canadian Class I device licence exist, and does it help in the United States?
There is no device licence for Class I in Canada. Section 26 of the Medical Devices Regulations requires a licence for Class II, III and IV devices only. United States classification is a separate exercise run by FDA against its own regulations, so a Canadian Class I designation carries no status across the border.
Is an MDEL the same thing as FDA establishment registration?
No. A medical device establishment licence is issued by Health Canada for the low-risk manufacturing class and to importers and distributors of all classes, so that they may import or distribute a device in Canada. FDA establishment registration is a separate annual filing in FDA's electronic system by the establishment that makes the device for the American market.
Do we need a United States agent if we sell through a US distributor?
Yes, if a foreign establishment is required to register under 21 CFR 807.40. The regulation requires the name, address and phone number of one United States agent as part of registration, and only one agent may be designated. The agent is not the same role as a distributor, although one company can hold both roles, and changes must be reported to FDA within 10 business days.
Are sterilization pouches and biological indicators covered by the sterilizer's clearance?
No. Under 21 CFR 880.6850 a sterilization wrap includes a pack, wrapper, bag or accessory intended to enclose a device being sterilized, and it is a Class II device. Biological and physical or chemical sterilization process indicators sit under 21 CFR 880.2800, also Class II. Each needs its own entry in the listing and its own answer on premarket notification.
Does the Quality Management System Regulation apply to a foreign accessory manufacturer?
Yes. Part 820 applies to finished devices imported or offered for import, and accessories are finished devices. Since 2 February 2026 the Quality Management System Regulation requires a quality management system that complies with ISO 13485:2016 as incorporated by reference, plus the additional requirements set out in 21 CFR 820.10.
Does a Canadian export certificate replace any American filing?
No. Part 4 of the Medical Devices Regulations provides for an export certificate in the form set out in Schedule 3, which some destination countries ask for as proof of the Canadian regulatory status of the device. FDA does not accept it in place of establishment registration or device listing.
Sources
CliniEco Medical distributes clinic, laboratory and care-home consumables in Canada under Health Canada MDEL #35334. Requirements that bind a particular practice remain the ones in its own written procedures and in the rules of the province where it operates.